Instrument Response
Use the Instrument Response option (UCERESPONSE GX) to calculate the instrument response (either static or dynamic) over known targets.
Instrument Response Test dialog options
Type of response test | Select the type of the response test, dynamic or static Script Parameter: UCERESPONSE.TYPE |
Application Notes
Static (bench) tests measure the sum of instrument and ambient noise (and also instrument drift). Dynamic tests along an actual survey line are necessary in order to measure motional and terrain noise. Emplacement of one or more calibration objects permits evaluation of amplitude response. A repeated dynamic test survey line also measures positional variation.
Please refer to the Help documents for the static (UCERESPONSES GX) and dynamic response (UCERESPONSED GX) tests.